44
Culligan® Series M2 Reverse Osmosis
44
Cat. No. 01023095
TDS Programming
If the measured TDS levels exceed the maximum allowed in the product water (see
“Table 6. Temperature Correction
), the RO controller displays an error condition (see
).
RUNNING
JAN-01-12 12:01P
1. From the default home screen, press
. The screen displays the main menu.
1) GO TO RUNNING
2) GO TO OFFLINE
3) INFORMATION
>4) SETUP
2. Press
to scroll to 4) SETUP.
3. Press
to select the SETUP menu.
2) LANGUAGE
3) POWERON MODE
4) FLUSH MODES
>5) ACCESSORIES
4. Press
to select 5)ACCESSORIES. The screen displays the ac-
cessories menu.
3) FLOW METERS
4) SWITCH INPTS
5) PRESS GAUGES
>6) TDS
5. Press
to select 6)TDS. The screen displays the TDS
settings.
TDS Probe Settings
Screen Display
Range
Changing the Setting
TDS PROBE:
NONE
None
Product
Both
For the M2 RO, there is only one TDS probe, so the correct set-
ting is PRODUCT. Press
or
and then
to change the
state if necessary.
ACTUAL TDS OUT
0
PPM
0–6000 (TDS1)
0–200 (TDS2)
Disconnect the TDS probe from the RO plumbing and submerge it
into a solution of known TDS. Press
or
to enter the actual TDS
value.
TEMP OFFSET
0 F
-50–50
Use this value to make minor corrections to the displayed temperature
value. For example if you use a handheld temperature device that
shows the water temperature two degrees warmer, change this value
to 2 to compensate.
TEMP COMP VALUE:
0.025
N/A
Please do not change this value unless you are told to do so by Cul-
ligan Rosemont. This value changes the slope used to interpret the
temperature.
RUNNING
JAN-01-12 12:01P
6. Press
to save the settings and return to the home screen.