Operation / Technical Manual
NORM-O-TEMP, Model 111Z
Page 62 of 81
6-16.
Electrical Safety
The NORM-O-TEMP
®
unit should be checked quarterly for electrical safety.
It is appropriate to perform such inspection upon receipt of a new unit prior to
acceptance, following repair and/or service of the unit, and then again during each
periodic inspection.
6-16.1.
Ground Integrity
Ground path resistance must be less than 0.2 ohms.
Using a Digital Multimeter or equivalent, verify resistance between the following
test points:
1.
between any exposed, unpainted part of the metal case of the NORM-O-
TEMP
®
and the ground conductor of the power plug,
2.
between the ground conductor and the quick-disconnect fittings, and
3.
between the ground conductor and the handle.
If a NORM-O-TEMP
®
unit has ground path resistance that exceeds the
appropriate limit, the cause should be investigated and corrected prior to
returning the unit to active service.
6-16.2.
Leakage Current
Leakage current must be less than 300 microamps.
Using a Safety Analyzer, verify leakage current under the following conditions
first while the unit is
grounded
and then again when the unit is
ungrounded
:
1.
Power
ON
, polarity normal and then polarity reverse. When measuring in
these conditions, measure when the NORM-O-TEMP
®
unit is heating and
circulating water (both the heater and pump are ON).
2.
Power
OFF
, polarity normal and then polarity reverse. When measuring
in these conditions, measure when the NORM-O-TEMP
®
unit is going
through self-calibration. This is when it is first turned on (both the heater
and pump are OFF).
If a NORM-O-TEMP
®
unit has leakage current that exceeds the appropriate limit
or has a significant increase in leakage current, the cause should be investigated
and corrected prior to returning the unit to active service.
Excessive leakage current is most commonly caused by a defective heater, but
other components can also fail in such a way as to increase leakage current. To
find the source of excessive leakage current, suspect components should be
sequentially disconnected until the source of the high leakage current is isolated.
Содержание NORM-O-TEP 111Z
Страница 14: ...Operation Technical Manual NORM O TEMP Model 111Z Page 14 of 81 Figure 1 NORM O TEMP unit Front View ...
Страница 16: ...Operation Technical Manual NORM O TEMP Model 111Z Page 16 of 81 Figure 2 NORM O TEMP unit Left Side View ...
Страница 18: ...Operation Technical Manual NORM O TEMP Model 111Z Page 18 of 81 Figure 3 NORM O TEMP unit Rear View ...
Страница 20: ...Operation Technical Manual NORM O TEMP Model 111Z Page 20 of 81 Figure 4 NORM O TEMP unit Right Side View ...
Страница 22: ...Operation Technical Manual NORM O TEMP Model 111Z Page 22 of 81 Figure 5 NORM O TEMP unit Model 111Z Top View ...