Operation / Technical Manual
NORM-O-TEMP, Model 111Z
Page 50 of 81
6-1.2.
Replace / Reinstall the Top of the Unit
A.
Re-Install top as it was removed in Section 6-1.1.
NOTE:
When securing top, be sure the fill spout gasket inner lip on
aluminum cover top is in the up position and fill spout goes directly into it.
6-1.3.
Removing the Three-Sided Enclosure Panel
A.
Loosen the top of the unit as described in Section 6-1.1.
B.
If an I.V. pole is attached to the stand, remove the two phillips screws from
the U-shaped bracket that secures the I.V. pole to the unit. Put the bracket
and screws aside.
C.
Remove the six remaining phillips head screws that secure the three-sided
panel to the unit.
D.
Remove the enclosure panel and put aside or discard if it is being replaced.
6-1.4.
Replace / Reinstall the Three-Sided Enclosure Panel
A.
Re-Install enclosure as it was removed in Section 6-1.3.
6-2.
Replacement of the Heater
A.
Obtain a replacement heater.
B.
Drain the reservoir and disconnect the power cord as described in Section 5-1.4.
C.
Remove the three-sided enclosure panel as described in Section 6-1.3.
D.
Locate the heater inserted in the lower front part of the reservoir.
E.
Disconnect the two heater wires attached to the power supply board.
F.
Using 5/16 inch nut driver unscrew the 6-32 kep nut to disconnect ground wire.
G.
Using a ⅞ inch wrench, unscrew the heater and pull it from the reservoir.
H.
Apply three (3) wraps of teflon tape around the threads of replacement heater.
I.
Insert and tighten (hand tight plus half turn) replacement heater into reservoir.
J.
Connect the ground wire to the grounding stud located on the reservoir frame
using a 6-32 kep nut.
K.
Connect both heater wires to the power supply board.
L.
Secure the three-sided enclosure panel as described in Section 6-1.4.
M.
Refill the reservoir as described in Section 5-1.5.
N.
Perform Electrical Safety Checks as described in Section 6-16. Perform both
Ground Integrity (Section 6-16.1) and Leakage Current (Section 6-16.2).
Содержание NORM-O-TEP 111Z
Страница 14: ...Operation Technical Manual NORM O TEMP Model 111Z Page 14 of 81 Figure 1 NORM O TEMP unit Front View ...
Страница 16: ...Operation Technical Manual NORM O TEMP Model 111Z Page 16 of 81 Figure 2 NORM O TEMP unit Left Side View ...
Страница 18: ...Operation Technical Manual NORM O TEMP Model 111Z Page 18 of 81 Figure 3 NORM O TEMP unit Rear View ...
Страница 20: ...Operation Technical Manual NORM O TEMP Model 111Z Page 20 of 81 Figure 4 NORM O TEMP unit Right Side View ...
Страница 22: ...Operation Technical Manual NORM O TEMP Model 111Z Page 22 of 81 Figure 5 NORM O TEMP unit Model 111Z Top View ...