Operation / Technical Manual
NORM-O-TEMP, Model 111Z
Page 45 of 81
After draining the reservoir as described in Section 5-1.4:
A.
Disconnect the unit from its power source.
B.
Remove the three sided enclosure panel.
C.
Locate the water filter assembly.
D.
Firmly grasp the notched rim of the cap of the assembly and turn it
counterclockwise. The cap may contain a small amount of water.
E.
Remove the wire mesh.
F.
Clean the wire mesh and the plastic cap. Be careful not to lose the black O-ring in
the rim of the plastic cap.
G.
Replace the wire mesh in the plastic cap and position the wire mesh, O-ring and
plastic cap.
H.
Screw the cap clockwise onto the fitting until it is secure.
I.
Replace the three sided enclosure panel.
J.
Replenish the water reservoir as described in Section 5-1.5.
WARNING
•
The repair, calibration, and servicing of the NORM-O-TEMP
®
should be performed by
qualified Medical Equipment Service Technicians, Certified Biomedical Electronics
Technicians, or Certified Clinical Engineers familiar with good repair practices for
servicing medical devices, and in accordance with instructions contained in this
manual.
Improper repair can result in damage to the NORM-O-TEMP
®
system
and patient injury.
•
Always
unplug the unit
before accessing internal components during service.
Failure to unplug the unit could result in
electric shock.
Содержание NORM-O-TEP 111Z
Страница 14: ...Operation Technical Manual NORM O TEMP Model 111Z Page 14 of 81 Figure 1 NORM O TEMP unit Front View ...
Страница 16: ...Operation Technical Manual NORM O TEMP Model 111Z Page 16 of 81 Figure 2 NORM O TEMP unit Left Side View ...
Страница 18: ...Operation Technical Manual NORM O TEMP Model 111Z Page 18 of 81 Figure 3 NORM O TEMP unit Rear View ...
Страница 20: ...Operation Technical Manual NORM O TEMP Model 111Z Page 20 of 81 Figure 4 NORM O TEMP unit Right Side View ...
Страница 22: ...Operation Technical Manual NORM O TEMP Model 111Z Page 22 of 81 Figure 5 NORM O TEMP unit Model 111Z Top View ...