Operation / Technical Manual
NORM-O-TEMP, Model 111Z
Page 44 of 81
5-1.5.
Replenishing the Reservoir / Fixing a Low Water Alarm
A.
Check that the drain hose is disconnected.
B.
Slide the water fill opening lid to the left
and gradually pour approximately 1½
gallons (5.7 liters) of distilled water with approximately 340 cc of propylene glycol
into the reservoir (DO NOT USE ALCOHOL, TAP WATER, OR DE-IONIZED
WATER).
C.
Stop pouring when the water reaches the strainer visible at the bottom of the
water fill opening.
D.
Proceed with normal operations; always check the water level before starting.
5-2.
Maintenance of the Water Filter
The NORM-O-TEMP
®
circulating system includes a water filter designed to clear the line
of any particulate matter as the water is pumped through the system. The water filter
assembly is a plastic T-shaped fitting that intersects the hose from the water flow
indicator to the reservoir. A stainless steel wire-mesh screen is located inside the clear
plastic cap of the water filter assembly.
Quarterly, or more often if deemed necessary, the water filter should be disassembled
and cleaned. To do so, the reservoir must first be drained and the three sided enclosure
panel removed. The cap of the water filter assembly is unscrewed. The wire mesh and
the plastic cap are cleaned. The parts are then reassembled.
CAUTION
•
Use distilled water only. Do Not Use De-Ionized water.
•
Do not
use alcohol. Alcohol may cause blanket/pad and unit deterioration.
•
Do not
overfill. Overfilling may result in overflow when the water in the blanket/pad
drains back into the system when the system is turned off.
CAUTION
•
Always drain the NORM-O-TEMP
®
to a sanitary drain because bacteria may be
present in the unit’s water supply.
Содержание NORM-O-TEP 111Z
Страница 14: ...Operation Technical Manual NORM O TEMP Model 111Z Page 14 of 81 Figure 1 NORM O TEMP unit Front View ...
Страница 16: ...Operation Technical Manual NORM O TEMP Model 111Z Page 16 of 81 Figure 2 NORM O TEMP unit Left Side View ...
Страница 18: ...Operation Technical Manual NORM O TEMP Model 111Z Page 18 of 81 Figure 3 NORM O TEMP unit Rear View ...
Страница 20: ...Operation Technical Manual NORM O TEMP Model 111Z Page 20 of 81 Figure 4 NORM O TEMP unit Right Side View ...
Страница 22: ...Operation Technical Manual NORM O TEMP Model 111Z Page 22 of 81 Figure 5 NORM O TEMP unit Model 111Z Top View ...