Operation / Technical Manual
NORM-O-TEMP, Model 111Z
Page 40 of 81
C.
Fill the reservoir with the appropriate amount of household bleach per chart
below.
UNIT
BLEACH
WATER
NORM-O-TEMP
®
, 111Z
168 Milliliters
5.4 Liters
D.
Fill the reservoir with warm distilled water.
E.
Turn unit on and circulate per the Chemical Cleaning Circulation Chart (above).
F.
Drain the unit as instructed in Step A.
G.
Rinse the unit three (3) times as described in Steps D-F. Be sure to omit the
household bleach (sodium hypochlorite).
H.
After the third rinse, drain the unit and fill the reservoir with distilled water and
circulate.
I.
Check the water with pH strips or other appropriate test method for detecting
bleach. If bleach is detected, repeat Steps F-H.
J.
If unit is being placed in dry storage skip steps K and L, continue with procedure
at step M.
K.
If unit is to be
returned to service, once no bleach is detected, add the
appropriate amount of U.S.P. Grade Propylene Glycol to the water reservoir per
the following chart:
UNIT
U.S.P. Grade Propylene Glycol Per Unit
NORM-O-TEMP
®
, 111Z
12 ounces (340cc)
L.
Continue to fill the water reservoir with distilled water. Then continue
procedure at Step O.
M.
When all fluid has been removed from the unit, disconnect the drain hose(s) and
wipe unit clean.
N.
Unit is now ready for storage.
When you are ready to return unit to service, replenish the reservoir(s) with
distilled water and 340 cc propylene glycol. Perform all the checks as described
in the respective Manual.
O.
Document unit maintenance.
Содержание NORM-O-TEP 111Z
Страница 14: ...Operation Technical Manual NORM O TEMP Model 111Z Page 14 of 81 Figure 1 NORM O TEMP unit Front View ...
Страница 16: ...Operation Technical Manual NORM O TEMP Model 111Z Page 16 of 81 Figure 2 NORM O TEMP unit Left Side View ...
Страница 18: ...Operation Technical Manual NORM O TEMP Model 111Z Page 18 of 81 Figure 3 NORM O TEMP unit Rear View ...
Страница 20: ...Operation Technical Manual NORM O TEMP Model 111Z Page 20 of 81 Figure 4 NORM O TEMP unit Right Side View ...
Страница 22: ...Operation Technical Manual NORM O TEMP Model 111Z Page 22 of 81 Figure 5 NORM O TEMP unit Model 111Z Top View ...