1-3.2.
External Features and Descriptions – Left Side View
The external features in Figure 2 of the NORM-O-TEMP
®
unit are described as
follows:
A.
The water flow indicator is a paddle wheel immersed in the path of the
circulating water with a window to the outside. As water is circulated
through the system, it must pass over the paddle wheel causing it to spin
(like a pinwheel). The water flow indicator provides a visual display of the
general rate at which the water is circulating. For example, if the unit is
circulating water but the connecting hose is pinched, the circulation of the
water is restricted. The change in water flow decreases the speed of the
paddle wheel.
The water flow indicator only spins when a blanket/pad or by-pass hose is
connected to the unit. It will not spin when water is circulated internally to
pre-condition the water. A total obstruction of the water path will cause the
paddle wheel to stop completely.
B.
Two screws on the left and right side of the unit secures the top to the base.
C.
Two female fittings, quick-disconnect return couplings, on the top row are
designed for water to flow in when the male coupling of the connecting hose
is attached.
D.
Two male fittings, quick-disconnect outlet couplings, on the bottom row are
designed for water to flow out when the female coupling of the connecting
hose is attached.
E.
The recessed handle, one on each side, is provided to assist in lifting the unit
on and off the I.V. stand or when carrying the unit.
Содержание NORM-O-TEP 111Z
Страница 14: ...Operation Technical Manual NORM O TEMP Model 111Z Page 14 of 81 Figure 1 NORM O TEMP unit Front View ...
Страница 16: ...Operation Technical Manual NORM O TEMP Model 111Z Page 16 of 81 Figure 2 NORM O TEMP unit Left Side View ...
Страница 18: ...Operation Technical Manual NORM O TEMP Model 111Z Page 18 of 81 Figure 3 NORM O TEMP unit Rear View ...
Страница 20: ...Operation Technical Manual NORM O TEMP Model 111Z Page 20 of 81 Figure 4 NORM O TEMP unit Right Side View ...
Страница 22: ...Operation Technical Manual NORM O TEMP Model 111Z Page 22 of 81 Figure 5 NORM O TEMP unit Model 111Z Top View ...