5-1.
Maintenance of Water Reservoir
The dual compartment reservoir holds approximately 1½ gallons (5.7 liters) of distilled
water that remains in the unit between periods of use.
Once a month the water reservoir should be drained and replenished. The drain hose,
included in the packaging of the unit, is attached and the water is pumped out of the unit.
When the LO WATER alarm sounds, the remaining water is drained by gravity.
Distilled/sterile water (or distilled water-cleaning agent preparation) is then poured into
the reservoir.
Quarterly, at a minimum, the Internal Cleaning / Dry Storage Procedure, listed below,
should be conducted.
NOTE: DO NOT USE DE-IONIZED WATER OR ALCOHOL.
Chemical Cleaning Circulation Chart
UNIT
MODE
TEMPERATURE
DURATION
NORM-O-TEMP
®
111Z
Heating
38°C
5 Minutes
5-1.1.
Internal Cleaning using Bleach / Dry Storage Procedure
For circulating water units utilizing household bleach
.
The following procedure cleans the fluid circuit in these products. The required
tools/supplies are appropriate drain hose(s) and/or hose connector(s), household
bleach, distilled water, U.S.P. Grade Propylene Glycol, and the appropriate AC
electrical power
Procedure:
A.
Drain the water from the reservoir as described in Section 5-1.4.
B.
Turn unit off (“O” position).
CAUTION
For safe handling and use of chemicals follow manufacturer guidelines.
Always drain the NORM-O-TEMP
®
to a sanitary drain because bacteria may be present
in the unit’s water supply.
Содержание NORM-O-TEP 111Z
Страница 14: ...Operation Technical Manual NORM O TEMP Model 111Z Page 14 of 81 Figure 1 NORM O TEMP unit Front View ...
Страница 16: ...Operation Technical Manual NORM O TEMP Model 111Z Page 16 of 81 Figure 2 NORM O TEMP unit Left Side View ...
Страница 18: ...Operation Technical Manual NORM O TEMP Model 111Z Page 18 of 81 Figure 3 NORM O TEMP unit Rear View ...
Страница 20: ...Operation Technical Manual NORM O TEMP Model 111Z Page 20 of 81 Figure 4 NORM O TEMP unit Right Side View ...
Страница 22: ...Operation Technical Manual NORM O TEMP Model 111Z Page 22 of 81 Figure 5 NORM O TEMP unit Model 111Z Top View ...