EZT-550 Controller Manual
61
Operand #
Description
Data Range
0031
Loop 2 (humidity) Percentage of Output
Read Only
Range = -100.0 to 100.0%
Note:
The value is read in whole numbers.
The decimal point is implied as the first digit in
the data value
Example:
Data value read of 379 = 37.9%
0054
Chamber Status
Indicates chamber trouble.
Read Only
1 = Product safety tripped
8 = Chamber high limit tripped
16 = Redundant product safety tripped
(when optional limit installed)
0055
Refrigeration System 1 Status
Indicates compressor trouble.
Read Only
1 = Pressure switch safety tripped
4 = Discharge temperature thermostat tripped
0056
Refrigeration System 2 Status
Indicates compressor trouble.
Read Only
1 = Pressure switch safety tripped
4 = Discharge temperature thermostat tripped
0057
Humidity System Status
Indicates humidity system trouble.
Read Only
Bit 0 = Humidity system disable
(air temperature out of range)
Bit 1 = Boiler over temperature
Bit 3 = Boiler water level low
Bit 4 = Reservoir level low
(when optional reservoir installed)
Note:
The alerts are bit oriented in the operand.
The actual value of the operand is dependent
upon the requests set.
0059
Maintenance Interval Alert
Indicates when a particular system needs
service.
Read Only
Bit 0 = Air circulator service request
Bit 1 = Refrigeration system service request
Bit 3 = Heater service request
Bit 4 = Humidity system service request
Bit 5 = Dry air purge system request
Bit 6 = Auxiliary cooling system service request
Note:
The alerts are bit oriented in the operand.
The actual value of the operand is dependent
upon the requests set.