EZT-550 Controller Manual
42
CAUTION
When changing display units, profile setpoints will not be converted between units.
Separate profiles are required for each temperature range. Equipment damage may
result if profiles of alternate units are used.
7.2.2 Maintenance Counters
The maintenance counters provide service interval alerts for chamber components. When a component exceeds
its service interval, a notification alarm is generated. In order to clear the alarm from the system alarm screen,
the maintenance counter must be cleared. To access the maintenance counters, select Maintenance Counters
from the setup menu.
The maintenance counters provide hours of operation monitoring for items such as air circulators and
compressors as well as cycle counts for solenoids, number of starts for compressors, etc. The service interval is
fixed at the factory for each component based on the required service interval. To clear a counter, select the
counter by touching the related component name. A message box will appear to verify the clear operation.
Select “Yes” to clear the counter and “No” to cancel.
The maintenance counters minimize unnecessary service by monitoring actual usage of the equipment rather
than using a fixed service interval. See your chamber manual for information regarding the types of service to be
performed for the different system components.