How to Calibrate Analog Inputs 1 and 2 (Air and DUT Temperature)
1) Select Analog Input 1 or Analog Input 2 from the Calibration menu. The calibration window for the input
will appear showing the minimum and maximum range, and the current value of the input.
2) Disconnect the sensor leads from the analog input on analog input card 1. The input card is connected
to the CPU module on the electrical sub-panel located beneath the chamber door behind the front
cover.
Note:
Analog input/output card 2 and the digital output card may not be present on all chambers.
They are installed based on options provided with the chamber.
3) Connect the leads from the calibration source to the input terminals.
Communication
Module
CPU Module
Analog Input/Output
Card 1
Analog Input/Output
Card 2
Options Digital
Output Card