EZT-550 Controller Manual
19
4) Input the required step data by touching the field entry boxes. Once touched, the keypad entry screen
will pop-up and allow the desired value to be entered.
Total Steps
The total steps entry is common to all steps of the profile. It specifies the total length of the profile. The
profile can be from 1 to 99 steps in length. This must be set to the correct number of steps in order for
the profile to work correctly.
Step Events
The step events button opens up the events list for the current step. From this window, select the
events that are to be on during the step. Once the selections are complete, return to the step entry
screen.
Note:
This must be done for each step individually. Also, for the chamber to operate, the relevant
chamber systems must be turned on via the event entry window. Even though a profile may
be in operation, if the events are not set, the chamber will not run.
TempSP
Target setpoint for air temperature during execution of step.
%RHSP
Target setpoint for humidity during execution of step.
Hrs: and Mins:
Time length of step in hours and minutes. Note that the time of a step can be zero (0). This is useful
for making instant setpoint changes between steps when a ramp is not desired.
Profile Recycle
The profile recycle allows the profile to jump between different steps within itself. This is useful for
repeating the same profile during extended cycle testing of product.
Recycle# - number of times to jump
Recycle Step – number of step to jump to once current step is complete