EZT-550 Controller Manual
33
3) Set the start and stop times and dates for the period in which the chamber is to operate.
Start/End (H:M) Start and stop times for the timer in hours and minutes. Note that he hours are
expressed in military time, i.e., 00-24 hours.
Start/End (M/D) Start and stop dates for the timer in months and days. Note that the year is not
required. The timer will start and stop based on a match of the month and day.
Note:
When the timer type is profile, the stop time and date does not have affect. Once a profile is
started, it will run until complete or stopped by the user.
4) Once all, settings have been made, press the Timer On/Off button. When the timer is off, the button will
read “Timer Off” and the green indicator light will be off. When the timer is enabled, the button will read
“Timer On” and the green indicator light will illuminate.
Note:
When the timer is enabled, the timer will assume control of the chamber from that point. The
user is locked out from making setpoint adjustments and running profiles. To regain control,
the timer must be turned off.
5.6 Defrost Settings
The defrost setting are for special systems with a hot-gas defrost option. The standard ZP(H) series chamber
does not utilize this feature. The menu item is therefore disabled.