EZT-550 Controller Manual
60
Operand #
Description
Data Range
0005
Humidity Setpoint
Used to set single setpoint humidity.
Read/Write
Limited to temperature range of chamber.
Typically –73°C (-100°F) to 190°C (374°F). See
chamber manual to verify valid safe operating
range.
Note:
The setpoint must be in whole numbers.
The decimal point is implied as the first digit in
the data value.
Example:
Setpoint of 42.4°C = 424 sent to operand
0016
Air Temperature
Read Only
Range = range of chamber
Note:
The process variable is read in whole
numbers. The decimal point is implied as the
first digit in the data value
Example:
Data value read of 997 = 99.7°
0017
DUT (Product) Temperature
Read Only
Range = range of chamber
Note:
The process variable is read in whole
numbers. The decimal point is implied as the
first digit in the data value
Example:
Data value read of 250 = 25.0°
0018
Relative Humidity
Read Only
Range = 0 – 100%
Note:
The process variable is read in whole
numbers. The decimal point is implied as the
first digit in the data value
Example:
Data value read of 500 = 50.0%
0029
Loop 1 (temperature) Percentage of Output
Read Only
Range = -100.0 to 100.0%
Note:
The value is read in whole numbers.
The decimal point is implied as the first digit in
the data value
Example:
Data value read of 379 = 37.9%