Sterilization
Chamber
Door open
Handle
Water Inlet Orifice
Temperature
Sensor Connection
Internal
Heating Element
Fig. 12
CRISTÓFOLI
A U T O C L
A V E S
I N S T R U C T I O N M A N U A L
Steam Internal Exit
Cover
Base
EXTERNAL STEAM EXIT
- Located in the back of the autoclave (Fig. 1, page 9), it has a diameter of 5/16” for the
connection of the discharge hose, which is then connected to the plumbing system or proper container for this
purpose. It releases the cold air in the beginning of the cycle and the steam at the end of it. See “
Installation
Instructions
” topic “
Hydraulic Installation
” (Page 9);
SOLENOID VALVES -
Internal components of the equipment responsible for the deareration, depressurization and
water supplying. The valves open the beginning of the heating stage eliminating the cold air from the chamber, then
it shuts down to allow pressure build up for the sterilization and opens again at the end of the cycle for the
depressurization of the chamber;
SOLENOID PUMP -
Internal device responsible for transporting the water from the reservoir to the chamber during the
automatic water supply;
Fig. 7
Fig. 8
Door gasket
Lid
Handle
Fig. 6
TRAYS HOLDER -
It's provided 1 tray holder for 4 trays (Fig. 11);
TRAYS -
Provided are 4 trays (Fig. 10), to keep the instruments to be sterilized free from any direct contact with the water
and internal surface of the autoclave chamber;
HEATING ELEMENT -
Electrical device responsible for heating the chamber (Fig. 9);
16
17
18
Heating Element Protector