![Cristofoli Autoclave Quadra 54 Скачать руководство пользователя страница 18](http://html1.mh-extra.com/html/cristofoli/autoclave-quadra-54/autoclave-quadra-54_manual_2686114018.webp)
d
c
2
17
Cancellation of the Cycle -
If the cycle is cancelled, for any reason, the display will alternate between two
messages, the message indicating the cycle cancellation (
CYCLE CANCELLED P:0.00 kgf/cm2
) and the message
indicating the possible cause for that, which can be:
lack of water in the RESERVoir
--- There’s no water in the reservoir or there was not water enough to fill the chamber
with the correct amount of water;
heating time limit
-------------------- The heating time limit was reached and there was no sterilization;
incompatible Temp. x PRESSURE
---- The relation between the temperature and pressure of the chamber is
;
incoherent
subPRESSure DETECTED
During the sterilization, the pressure decreased below the acceptable level
--------------
;
OverPRESSure DETECTED
During the sterilization, the pressure increased above the acceptable level
-------------
;
CANCELLED by user
The operator pressed the
CANCEL
key
-------------------
;
door open
------------------------------ The operator opened the door during the heating or drying stage (before the valve
shut down);
power outage
------------------------
.
There was a power outage at the workplace (during the sterilization stage only)
3
If it’s necessary to interrupt/cancel the heating, sterilization or drying cycle, just press the
CANCEL
key. In this case,
after the beep and the indication
cycle cancelled
on the display, the autoclave
will beep continuously until the
2
pressure falls to 0,0 kgf/cm and the
CANCEL
key is pressed again to confirm the cancellation of the cycle, there will
be no effect in case the
CANCEL
key is pressed while there’s still pressure in the chamber;
CRISTÓFOLI
A U T O C L
A V E S
I N S T R U C T I O N M A N U A L
DEPRESSURIZE?
ST : YES CAN : NO
Some situations may cause interruption and automatic cancellation of the cycle when:
the ideal pressure/temperature is not reached due to steam/pressure leak or overloaded chamber. In this case,
the cancellation will occur within 50 minutes at the most.
the sterilization cycle is initiated and the chamber is not supplied with water within 20 seconds, the cycle will be
canceled automatically. In case there’s a lack of water while filling up the chamber, the autoclave will show the
message
and beep until the
key is pressed to resume supply.
FILL UP RESERVOIR
Start
there’s a p
ower outage or voltage fluctuations. The depressurization will occur when power returns. See note
below;
t
urning the autoclave on, there’s already some pressure in the chamber. When that happens, the autoclave
will beep continuously, the display will show
cycle cancelled
and the autoclave will depressurize. See note below;
Confirm the cancellation of the cycle manually by pressing the
CANCEL
key;
Check the possible causes, take the necessary measures to correct the problem and perform a new cycle
to reprocess the articles according to the instructions on “
How to Use the Autoclave Quadra 54
” (Page 14).
Before starting a new cycle, t
he operator must check if there is water left in the chamber, which must be
removed manually through the door with a clean and dry cloth that does not shed or through the function
“
Chamber Water Draining
” as described on topic 5 (Page 18).
For your safety, remember to use
Attention!
PPE (Personal Protective Equipment, like proper latex gloves);
Note:
In the situations
c
and
d
above, if there’s pressure inside the chamber, the display will show the question:
By pressing the
START
key, the valve opens for depressurization; by pressing the
CANCEL
key,
the depressurization will happen
, with the valve clo
this will prevent the liquid
gradually
sed,
from boiling over inside the containers when the liquids sterilization cycle was selected;
ADVERSE SITUATIONS
a
b
1
Содержание Autoclave Quadra 54