Check if the power cable is plugged in, switch on the circuit breaker in the back of the autoclave, the display will
show the following mesages: and , next, the display will go out. In order to turn the
display on again, press the
START
key for 2 seconds, the display will show the initial information (brand, model and
software version), the autoclave will beep twice and the display will show
select program
.
bootloader V000
aguarde . . .
bootloader V000
executando . . .
HOW TO USE THE AUTOCLAVE QUADRA 54
a
SAFETY NOTES
b
Open the door of the autoclave and load it with the materials to be sterilized, be careful not to lean them against
the chamber walls or internal steam exits, that will cause interference in the cycle and damages to the articles.
Do
not overload the autoclave
.
Warning!
- During the autoclave operation, it is perfectly normal to hear some noises, some low, some loud. The
noises are generated by the valves opening and closing, deaeration and depressurization, they are part of the proper
functioning of the equipment. The door gasket and the VSPF valve, are safety mechanisms that when activated
discharge pressure automatically producing a loud noise. This autoclave must be installed in a proper and exclusive
sterilization room. Cristófoli is not responsible for accidents that might occur due to the starts caused by the noises
produced by the autoclave.
The symbol 14 appears in some places on the autoclave. This means those items require special attention and
that the user must observe their references in the Instruction Manual provided with the equipment regarding potential
hazards offered by them and any actions to be taken should an adverse situation occur.
Cristófoli Equipamentos de Biossegurança Ltda, does not take any responsibility for failures and/or accidents
caused by the non-observance of this warning.
ATTENTION!
Before starting a new sterilization cycle, make sure the reservoir has enough water to supply the autoclave need
(at least 1,5 liters).
Use only distilled water for the sterilization
(in accordance with the EN 13060)
.
Non-compliance with this
recommendation may cause obstructions of the autoclave hydraulic system (internal tubing and/or valves), stains on the
instruments and loss of warranty.
Close the door of the autoclave. To close it correctly, press it against the chamber, slide the handle downwards until
it reaches the end of the course (Fig. 6, page 11). To open the autoclave, slide the handle upwards and open the
door.
ATTENTION!
In case the door of the autoclave is not properly closed, the program will not start. The autoclave will beep
and the display will show the message:
close the door!