Nellcor™ Pulse Oximetry Sensor Connection
Operator’s Manual
3-5
Note:
Two types of battery are available: the standard 5-hour and optional 10-hour.
Note:
Even if the monitoring system is turned off, the
Battery Charge Indicator
remains lit while
the battery recharges.
Note:
A full charge of a depleted battery takes more than four (4) hours for a 5-hour battery or
eight (8) hours for a 10-hour battery.
Plug the monitoring system into an AC outlet to charge the battery for a minimum of three
(3) minutes prior to turning on any monitoring system with a completely discharged
battery. When operating on internal battery, the monitoring system battery status icon
indicates the battery charge condition.
To charge the internal battery:
1.
Connect the monitoring system to AC power to charge a low or depleted battery. Ref-
erence
2.
Verify the
AC Power Indicator
and
Battery Charge Indicator
light.
3.5
Nellcor™ Pulse Oximetry Sensor Connection
WARNING:
Incorrect application or use of an SpO
2
sensor can cause tissue damage. Do not wrap
the sensor too tightly, apply supplemental tape, or leave a sensor too long on one
place. Inspect the sensor site as directed in the
Directions for Use
to ensure skin
integrity, correct positioning, and adhesion of the sensor.
WARNING:
Do not use any other cables to extend the length of the Covidien-approved interface
cable. Increasing the length will degrade signal quality and may lead to inaccurate
measurements.
Содержание Nellcor Series
Страница 1: ...Operator s Manual Nellcor Bedside SpO2 Patient Monitoring System TM...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 12: ...Page Left Intentionally Blank x...
Страница 20: ...Introduction Page Left Intentionally Blank 1 8 Operator s Manual...
Страница 36: ...Installation Page Left Intentionally Blank 3 8 Operator s Manual...
Страница 118: ...Product Specifications Page Left Intentionally Blank 11 18 Operator s Manual...
Страница 139: ...Results Operator s Manual A 21 Figure A 17 Correlation Plot for SpO2 All Data No Motion SaO2 vs SpO2 1 SpO2 2 SaO2...
Страница 156: ...Clinical Studies Page Left Intentionally Blank A 38 Operator s Manual...
Страница 159: ......