Theory of Operations
10-8
Operator’s Manual
10.6.4
The SatSeconds™ Safety Net
The SatSeconds™ “Safety Net” is for patients with saturation levels frequently below
the limit, but not staying below the limit long enough for the SatSeconds™ time
setting to be reached. When three or more limit violations occur within 60 seconds,
an alarm sounds even if the SatSeconds™ time setting has not been reached.