Product Specifications
11-4
Operator’s Manual
11.6
Performance Specifications
Key Beep
Volume level
Adjustable (Off, level 1-7),
(Invalid key presses are ignored)
Pitch (± 20Hz)
440 Hz (valid), 168 Hz (invalid)
Pulse width (± 20msec)
110 msec
Number of pulses
N/A
Repetitions
No repeat
POST Pass Tone
Volume level
Not changeable
Pitch (± 20Hz)
780 Hz
Pulse width (± 20msec)
1500 msec
Number of pulses
N/A
Repetitions
No repeat
Table 11-2.
Tone Definitions (Continued)
Tone Category
Description
Table 11-3.
Trends
Types
Graphical and Tabular
Memory
Saves total 88000 data events
Saves date and time, alarm conditions,
pulse rate, and SpO
2
measurements
Graphical Format
Total 2 graphs
A graph for SpO
2
parameters
A graph for Pulse Rate parameters
Tabular Format
One table for all parameters