Installation
3-4
Operator’s Manual
3.4.2
Internal Battery Power
WARNING:
The amount of time between the low battery alarm and power off becomes shorter
as the battery accumulates charge/discharge cycles.
Note:
Remove the battery if the monitoring system is not likely to be used for six (6) months.
Note:
Covidien strongly recommends fully recharging the battery whenever the time between
recharges exceeds six (6) months.
Note:
The monitoring system may not operate if the battery charge is critically low.
Note:
Covidien strongly recommends keeping the monitoring system connected to AC power
during continuous operation or to recharge the internal battery.
The monitoring system has an internal battery that powers the monitoring system
when AC power is not available. The monitoring system cannot operate with a fully
discharged battery. A lit battery status icon indicates the monitoring system is
running on battery power.
Before using the internal battery, perform a safety check of the equipment. Refer-
ence
A new, fully charged battery will provide its optimal number of operational hours
under these normal conditions:
•
Operating in Normal mode (Measuring SpO
2
and PR with plethysmograph display)
•
Setting for pulse beep indicator is ON (pulse volume: 4 (Default))
•
Sleep Study mode is inactive
•
Setting for SatSeconds™ is ON
•
Experiencing no alarm condition
•
Operating at ambient temperature of 25°C (±5°C)
Содержание Nellcor Series
Страница 1: ...Operator s Manual Nellcor Bedside SpO2 Patient Monitoring System TM...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 12: ...Page Left Intentionally Blank x...
Страница 20: ...Introduction Page Left Intentionally Blank 1 8 Operator s Manual...
Страница 36: ...Installation Page Left Intentionally Blank 3 8 Operator s Manual...
Страница 118: ...Product Specifications Page Left Intentionally Blank 11 18 Operator s Manual...
Страница 139: ...Results Operator s Manual A 21 Figure A 17 Correlation Plot for SpO2 All Data No Motion SaO2 vs SpO2 1 SpO2 2 SaO2...
Страница 156: ...Clinical Studies Page Left Intentionally Blank A 38 Operator s Manual...
Страница 159: ......