Clinical Studies
A-4
Operator’s Manual
Modified Bland-Altman plots for all the data are presented in Figure A-1 and Figure
A-2 for SpO
2
and pulse rate respectively.
Note:
Each individual subject is represented by a unique color on the plots. Subject identification
numbers are indicated in the legend to the left of each plot.
Table A-2.
Pulse Rate Accuracy Results (No Motion)
Sensor
Motion
# of
Data
Points
A
RMS
(BPM)
Pulse Rate
Acceptance
Criteria
(BPM)
Pass/Fail
MAXA
No
1154
0.76
3.0
Pass
MAXN
No
874
0.74
3.0
Pass
MAXFAST
No
281
0.81
3.0
Pass
SC-A
No
636
2.20
3.0
Pass
DS-100A
No
444
0.77
3.0
Pass
OxiCliq-A
No
499
0.79
3.0
Pass
D-YSE
No
473
0.98
3.0
Pass