External Data Communication
Operator’s Manual
5-5
When the nurse call polarity is set to NORMALLY –, the nurse call relay operation is
as follows:
Pin 2 is a common lead for both relays.
5.4.2
Trend Data Download
Caution:
Anyone who connects a PC to the data output port configures a medical system and
is therefore responsible for ensuring that the system complies with the requirements
of IEC 60601-1-1 and the electromagnetic compatibility IEC 60601-1-2.
Caution:
Patient motion, secondary to a variety of external factors, may compromise the
presence or accuracy of the displayed values.
Connect the mini-USB port to a PC for downloading trend data. Any PC connected
to the data port must be certified according to IEC 60950. All combinations of equip-
ment must be in compliance with IEC 60601-1-1 system requirements. Use either
ASCII communication protocol.
•
Nellcor™ ASCII protocol (ASCII 1)
Table 5-1.
Nurse Call Relay Pins States for NO
Monitoring system ON
Monitoring system OFF
NO
No alarm
or alarm silenced
Audible alarm
Pin 1 and Pin 2
Open
Closed
Closed
Pin 2 and Pin 3
Closed
Open
Open
Table 5-2.
Nurse Call Relay Pins States for NORMALLY –
Monitoring system ON
Monitoring system OFF
NORMALLY –
No alarm
or alarm silenced
Audible alarm
Pin 1 and Pin 2
Closed
Open
Closed
Pin 2 and Pin 3
Open
Closed
Open
Содержание Nellcor Series
Страница 1: ...Operator s Manual Nellcor Bedside SpO2 Patient Monitoring System TM...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 12: ...Page Left Intentionally Blank x...
Страница 20: ...Introduction Page Left Intentionally Blank 1 8 Operator s Manual...
Страница 36: ...Installation Page Left Intentionally Blank 3 8 Operator s Manual...
Страница 118: ...Product Specifications Page Left Intentionally Blank 11 18 Operator s Manual...
Страница 139: ...Results Operator s Manual A 21 Figure A 17 Correlation Plot for SpO2 All Data No Motion SaO2 vs SpO2 1 SpO2 2 SaO2...
Страница 156: ...Clinical Studies Page Left Intentionally Blank A 38 Operator s Manual...
Страница 159: ......