Operation
4-10
Operator’s Manual
4.6
Manage Alarms and Alarm Limits
WARNING:
Setting alarm limits to off or extreme high or low values will reduce alarm efficacy.
WARNING:
Do not silence the audible alarm or decrease its volume if patient safety could be
compromised.
WARNING:
Prior to each use of the monitoring system, check the alarm limits to ensure they are
appropriate for the patient being monitored. Ensure alarm limits do not exceed the
standard thresholds set by the institution.
WARNING:
Ensure the speaker is clear of any obstruction. Failure to do so could result in an
inaudible alarm tone.
When the monitoring system detects certain conditions that require user attention,
the monitoring system enters an alarm state.
The monitoring system uses both visual and audible indicators to identify high-prior-
ity, medium-priority, and low-priority alarms. Audible alarms include pitched tones,
beeps, and a buzzing tone. High priority alarms take precedence over medium- and
low-priority alarms. Reference
Table 4-1.
Alarm Conditions
Priority
Rate
Color
Messages
High
Sounds every 4
s
Red
Steady message,
Fast flashing numeric
SpO
2
Loss of Pulse
Critically Low-Battery condition
Medium
Sounds every 8 s
Yellow
Steady message,
Slow flashing numeric
High Pulse Rate limits violated
Low Pulse Rate limits violated
High SpO
2
limits violated
Low SpO
2
limits violated
Low
Sounds every 16 s
Steady yellow
SpO
2
Cable/Sensor Disconnect
SpO
2
Sensor Off
Low Battery
Technical System Error: EEE 001
Содержание Nellcor Series
Страница 1: ...Operator s Manual Nellcor Bedside SpO2 Patient Monitoring System TM...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 12: ...Page Left Intentionally Blank x...
Страница 20: ...Introduction Page Left Intentionally Blank 1 8 Operator s Manual...
Страница 36: ...Installation Page Left Intentionally Blank 3 8 Operator s Manual...
Страница 118: ...Product Specifications Page Left Intentionally Blank 11 18 Operator s Manual...
Страница 139: ...Results Operator s Manual A 21 Figure A 17 Correlation Plot for SpO2 All Data No Motion SaO2 vs SpO2 1 SpO2 2 SaO2...
Страница 156: ...Clinical Studies Page Left Intentionally Blank A 38 Operator s Manual...
Страница 159: ......