Preventive Maintenance
7-4
Operator’s Manual
7.5
Periodic Safety Checks
Covidien recommends a qualified service technician perform the following checks
every 24 months.
•
Inspect the equipment for mechanical and functional damage or deterioration.
•
Inspect the safety relevant labels for legibility. Contact Covidien or a local Covidien rep-
resentative, if labels are damaged or illegible.
•
Ensure all user interface keys, cables, and accessories function normally.
7.6
Service
WARNING:
Only a qualified service technician should remove the cover or access any internal
components.
Caution:
Dispose of monitoring system in accordance with local requirements and
regulations.
The monitoring system requires no routine service other than cleaning, battery
maintenance, and service activity mandated by the institution. For more informa-
tion, reference the
Service Manual
.
•
The monitoring system requires no calibration.
•
Have a qualified service technician replace the battery at least every two (2) years.
•
If service is necessary, contact Technical Services or a qualified service technician. Refer-
ence
Содержание Nellcor Series
Страница 1: ...Operator s Manual Nellcor Bedside SpO2 Patient Monitoring System TM...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 12: ...Page Left Intentionally Blank x...
Страница 20: ...Introduction Page Left Intentionally Blank 1 8 Operator s Manual...
Страница 36: ...Installation Page Left Intentionally Blank 3 8 Operator s Manual...
Страница 118: ...Product Specifications Page Left Intentionally Blank 11 18 Operator s Manual...
Страница 139: ...Results Operator s Manual A 21 Figure A 17 Correlation Plot for SpO2 All Data No Motion SaO2 vs SpO2 1 SpO2 2 SaO2...
Страница 156: ...Clinical Studies Page Left Intentionally Blank A 38 Operator s Manual...
Страница 159: ......