A1000 INSTALLATION, OPERATION, & MAINTENANCE MANUAL
February 2016 | 256574-ENG
R01 8-1
8 TECHNICAL
SPECIFICATIONS
Conviron maintains a policy of continual improvement and reserves the right to change the
technical characteristics of the A1000 without prior notice.
Specifications
A1000
PG Kit
AR Kit
TC Kit
IN Kit
Kit Crated Weight
(lbs./kg)
286/130 286/130 506/230 286/130
Chamber Crated
Weight (lbs./kg)
790 / 360
Exterior Dimensions
Height (in./mm)
79 / 2005
Width (in./mm)
41 / 1040
Depth (in./mm)
32.5 / 825
Interior Dimensions
Height (in./mm)
52.5 / 1334
Width (in./mm)
37 / 940
Depth (in./mm)
25 / 635
Power Requirements
60Hz
120VAC, 20A, 1
50Hz
220V / 240V, 16A, 1
Cooling Requirements
Direct Expansion
Air Cooled
40°F to 85°F (4.4°C to 29.4°C) at 150 psi maximum with10 psi pressure differential
Environmental Requirements
Temperature
Humidity
Altitude
68
F to 86
F (20
C to 30
C)
Up to 55% RH, non-condensing
7000 ft. (2134m) above sea level