Kit Installation
2-16 February 2016 | 256574-ENG
R01
2.3.2.5 Installing the Lamp Canopy Supports and Lamp Canopy
The lamp canopy location in the PG kit is predetermined by the position of the air intake and
cooling fan cavities in the side walls.
To install the lamp canopy supports
and lamp canopy:
1. Position the lamp canopy supports,
Item J, directly below the cavities in
each side wall (Figure 2-17).
a. Install the support with the flat side
up by hooking the upper tab into
the slots. The flat side has an oval
hole in it to locate the canopy in
position.
b. Press lightly on the bottom side,
snapping it into the corresponding
slot approximately 1” (25 mm)
below the top one.
2. Carefully slide the canopy, Item K,
onto the supports with the canopy
plug located on the right hand side of
the chamber.
Figure 2-17
PG Lamp Canopy Supports
Figure 2-18
PG Lamp Canopy Installed
Incorrect installation may result in user injury due to overheating of the canopy or
significantly reduced chamber performance. The canopy must be installed so the
intake and fan cavities align with the ports on the sides of the canopy.
Ensure at least 2” (50mm) of free space around the side walls of the chamber interior
to facilitate proper air circulation.
K
J
J