2.5.2.2 Installing the Lamp Canopies
The AR kit provides two adjustable lamp
canopies to accommodate the specific
requirements. However, there is a
limitation determined by the distance from
the top canopy to the fan housing. If the
canopy is installed too close, the airflow
may be significantly reduced, affecting the
unit’s performance.
The maximum growth height for the AR
kit is 18” (450mm). The growth height is
adjustable according to the requirements
of the experiment.
Another option is to have different growth
heights between the two tiers, or to install
a single canopy with one large growth
height. Figure 4 34 shows a chamber
configured for the maximum growth
height between canopies of 18” (450mm).
Figure 2-34
AR Shelf Supports Installed for Maximum
Growth Height
Do not install the top canopy closer than 3” (75mm) from the air vent located in the top
of the chamber.
Keep all unused receptacles capped to prevent an electric shock hazard.