A1000 INSTALLATION, OPERATION, & MAINTENANCE MANUAL
February 2016 | 256574-ENG
R01 2-5
2.2.2.2 Installing the Back Wall Plenum
The back wall plenum supports the rear
edge of the air shelves and closes the
front of the chamber plenum to distribute
air to the shelves.
To install the back wall plenum:
1. Remove the rubber instrumentation
port plug from the inside wall and
temporarily re-install it in the port on
the outside wall.
2. With the plenum handle at the top and
the four rows of air slots towards the
bottom, gently slide the back wall
plenum, Item B, into the chamber.
3. Position the plenum at a slight angle
to clear the protruding lamp
receptacles and the aspirator and
push it into place against the chamber
back wall.
4. Engage the locating pins above the
handle on the top of the plenum into
the holes in the top of the chamber.
5. Secure the plenum to the chamber
with twelve (12) screws, Item C
(Figure 2-3).
a. Install the eight (8) vertical screws
loosely to hold the plenum in
place.
b. Gently pry the back wall plenum
upwards to align the four (4)
bottom holes in the plenum with
the holes in the bottom support
and install the screws.
c. Tighten all screws to full
resistance.
Figure 2-3
TC Back Wall Plenum Installed
C
B
Aspirator
Rubber Plug
Lamp
Receptacles
Содержание ADAPTIS A1000
Страница 1: ...A1000 INSTALLATION OPERATION MAINTENANCE MANUAL 256574 ENG R01 ...
Страница 14: ...Precautions 1 4 February 2016 256574 ENG R01 ...
Страница 46: ...Kit Installation 2 32 February 2016 256574 ENG R01 ...
Страница 54: ...Chamber Start Up and Operation 5 4 February 2016 256574 ENG R01 ...
Страница 63: ......