414L-M-V3
4.3 DENSITY CORRECTION FOR GENERAL LIQUIDS
If density correction for general liquids is selected, a five point
temperature density table can be programmed. The instrument uses
this table to determine the density of the product after measuring the
temperature of the fluid.
For known fluids, the temperature-density characteristics are well
documented in standard chemical reference books. One to five correction
points can be programmed, and the user inputs the number of points during
programming.
The instrument will also allow a constant density value to be programmed
by selecting only one correction point. In this case, the temperature input
is ignored. This feature is useful if the temperature sensor is removed for
maintenance.
The flow equation for volumetric flowmeters is:
Mass Flow = R x Density
where R is the flowrate defined in sections 3.4.1 and 3.4.2.
Density is the density determined from the
temperature-density table.
When programming the temperature-density table, the density can
be programmed in any units, but the units must be consistent
with the flowrate R, and the required units for MASS flow.
For example, if R is in litres and the mass
is required in kilograms, then the density must obviously be programmed
as kg/litre.
36
Temperature Compensation