414L-M-V3
Temperature Correction
Density Correction for
General Liquids:
A five point temperature density table can
be programmed.
Temperature Range:
-273°C to 200°C.
(-459°F to 392°F).
LPG:
To API/ASTM Tables 34 and 54
Accuracy:
0.04% for metric.
0.12% for US.
Temperature Range:
-45°C to 60°C.
(-49°F to 140°F).
Volume Correction for
Petroleums:
Correction to API/ASTM Tables as follows;
Crude:
Tables 24A and 54A.
Gas, Jet, Oils:
Tables 24B and 54B.
Lube Oil:
Tables 24D and 54D.
Accuracy:
0.075%
Temperature Range:
-100°C to 200°C.
(-148°F to 392°F).
Volume Correction for
General Liquids:
Correction to any base temperature using a
constant co-efficient of volume expansion.
Accuracy:
0.05%
Temperature Range:
-273°C to 200°C.
(-459°F to 392°F).
Density Meter Input
A density meter with 4-20mA output can
be connected, in place of a temperature
sensor, to provide a mass flow reading.
(Model 414LA only.)
8
Specification