414L-M-V3
3.4 CALCULATION OF RATE AND TOTAL
3.4.1 Frequency Input
For instruments without temperature compensation, the volumetric
flowrate at flow conditions, R, is calculated as follows:
R =
f x H
S
where
f
is the input frequency in Hz.
H
is the timebase of rate and is 1 for seconds, 60 for
minutes, 3600 for hours and 86,400 for days.
S
is the Scaling Factor.
The Scaling Factor,
S
, is equal to the K-factor of the flowmeter expressed
in pulses per unit volume.
The flowrate,
R
, as expressed in the above equation, is the volumetric
flowrate at flow conditions. With temperature compensation, the flowrate, R,
is corrected to give the flowrate at a base temperature, or the mass flow.
Temperature Correction is covered in detail in Section 4.
The user programs the Scaling Factor and selects the timebase during
the Calibration procedure as detailed in Section 6 of this manual.
When non-linearity correction is programmed, up to 10 scaling factors
are programmed to cover different frequency ranges. The instrument
will then automatically select the correct scaling factor to be applied at
the measured frequency.
Operation
23