Date
1.
Press
SET
until the display shows [SEt dAtE] and press
CAL
.
2.
The display shows [dAtE=on] or [dAtE=OFF]. Select the desired mode and press
CAL
.
3.
Program the date (respectively year, month and day) by pressing
CAL
each
time.
Time
1.
Press
SET
until the display shows [SEt tiME] and press
CAL
.
2.
The display shows [tiME=on] or [tiME=OFF]. Select the desired mode and press
CAL
.
3.
Program the time (respectively hour, minutes and seconds) by pressing
CAL
each time.
Temperature
1.
Press
SET
until the display shows [SEt °C] and press
CAL
.
2.
The display will show the measured temperature e.g. [°C=23.7][Aut]. If neces-
sary, calibrate your probe and press
CAL
. Press
RES
to reset the temperature
calibration to the original factory calibration.
3.
When no Pt1000 temperature probe is connected to the instrument the display
will show the manual temperature e.g. [°C=23.7][MAn]. Shift to the proper
temperature and press
CAL
.
Recorder
1.
Press
SET
until the display shows [SEt rEC] and press
CAL
.
2.
The display shows for all channels e.g. [rEc=PH], [rEc=MV], [rEc=EC],
[rEc=°C]...etc. Select the desired mode and press
CAL
.
3.
This step is
skipped
unless [rEc=EC] has been chosen. Press
CAL
and the display
shows the conductivity measuring range and units, e.g. [rEc 0.01MS][Unit].
Select the desired range and press
CAL
.
4.
The display shows e.g. [rEc=6.00][LoW]. Select the desired minimum level cor-
responding to a 0 V recorder output and press
CAL
.
5.
The display shows e.g. [rEc=8.00][HiGH]. Select the desired maximum level
corresponding to a 2 V recorder output and press
CAL
.
Channels
1.
Press
SET
until the display shows [SEt CHAn] and press
CAL
.
2.
The display shows e.g. [PH=on] or [PH=OFF]...etc. Select the desired mode
(respectively pH, mV and EC) by pressing
CAL
each time.
3.
The display shows e.g. [rEF=25°C]. Select the temperature to which all future
conductivity measurements will be referred and press
CAL
.