R735 • www.consort.be
11 / 36
Data-logging
Pre-set and start the data-logging:
1.
Press
SET
until the display shows [SEt LOG] and press
CAL
.
2.
Select [ConFiG][LOG] and press
CAL
.
3.
Choose between [tiMEr], [MAnuAL] or [OFF]. Select the desired mode and press
CAL
.
4.
This step is
skipped
unless automatic data-logging is chosen (timer). The dis-
play shows e.g. [dt=100]. Select the desired time interval (0...9999 s) between
the automatic data-logging and press
CAL
.
5.
When manual data-logging has been selected, press
to put a next meas-
urement into memory. In the meantime the display shows the storage number
e.g. [LOG 7].
View the stored values on the display:
1.
Press
SET
until the display shows [SEt LOG] and press
CAL
.
2.
Select [LSt dAtA] and press
CAL
.
3.
The display shows the first data-logged value in memory (a blinking [n]
appears). Press
Ù
or
Ú
to view all other values.
4.
While pressing
HOLD
the corresponding date is shown.
5.
While pressing
RES
the corresponding time is shown.
6.
While pressing
CAL
the corresponding number is shown.
7.
Press
to send the recalled data to a printer.
8.
Press
SET
to return.
Erase the the stored values:
1.
Press
SET
until the display shows [SEt LOG] and press
CAL
.
2.
Select [CLr LOG] and press
CAL
.
Send the stored values to a printer or computer:
1.
Press
SET
until the display shows [SEt LOG] and press
CAL
.
2.
Select [Prt dAtA] and press
CAL
to print the data.
3.
The display shows [PrintinG] while all data-logged values, including corre-
sponding dates and times, are sent to the RS232 output in the following for-
mat, compatible with EXCEL™, LOTUS 123™,....The columns are separated by
a TAB character:
Identification No: 037
# VALUE UNIT C CH H D
0001 7.34 pH 23.7 2 13:19:57 29/01/95
0002 7.21 pH 23.7 2 13:20:57 29/01/95