Perform the following steps:
16.
reset the meter (switch on while holding MODE pressed).
yes?
15.
store all pH electrodes in a 3...4 KCl solution overnight.
yes?
18.
select the °C mode, measure temperature of solutions and read the display.
°C:
19.
select the mV mode (do
not
calibrate!).
yes?
20.
connect the pH electrode, rinse, dip in fresh buffer 7 (stir!) and read display.
mV:
17.
short-circuit the pH input, select the mV mode and read display.
mV:
21.
rinse the pH electrode, dip in fresh buffer 4 (stir!) and read display.
mV:
14.
type of connected electrode:
22.
connect a new pH electrode, rinse, dip in fresh buffer 7 (stir!) and read display.
mV:
23.
rinse this pH electrode, dip in fresh buffer 4 (stir!) and read display.
mV: