R735 • www.consort.be
36 / 36
Perform the following steps:
16.
reset the meter (switch on while holding MODE pressed).
yes?
15.
store all conductivity cells in distilled water with some detergent overnight.
yes?
18.
connect a 1000 ohm resistance to the conductivity input and read display.
µS:
19.
select the °C mode, measure temperature of solutions and read the display.
°C:
20.
rinse the conductivity cell with a 0.01 M KCl calibration solution (1413 µS).
yes?
17.
leave the EC input open, select the conductivity mode and read display.
µS:
21.
connect the conductivity cell, dip in fresh cal. solution (stir!) and read display.
µS:
14.
type of connected electrode:
22.
rinse a new conductivity cell with a 0.01 M KCl calibration solution (1413 µS).
yes?
23.
connect this conductivity cell, dip in fresh cal. solution (stir!) and read display.
µS: