EPIQ 5 & 7 Training Manual
© 2016 Conquest Imaging
84
Post VIO or Graphics Card Replacement
After you replace a video module VIO, SVIO etc. or a graphics card
you should perform the display assignment procedure so that the
touch screen will display correctly.
Display Assignment
The display assignment procedure is a script that identifies the two
displays to the ultrasound system’s operation system. The monitor is
identified as
Display A
and the touch screen is
Display B
. If you do not
run the script, the touch screen will not display properly and some or
all of the touch controls are not available.
To set the monitor and touch screen for proper display:
Press
Support
.
Click
Test/Utilities
Hardware Utilities
.
In the dialog box that reminds you to reboot the system when
you exit
SupportConnect
click
Yes
.
On the
Interactive Utilities
display, click
Display Assignment
Run Test
. A message indicates the display-assignment status.
What do they do if not OK?? What is the message?
Click
OK
.
iButton
The iButton is a component that contains the chip-ID serial number
associated with certain licensed options. The location of the iButton is
different for different versions of the EPIQ systems. When replacing a
board where the iButton resides you must remove it from the board
you are replacing and replace it on the new board in order to retain
these licensed options. The system reads the number through the
PCI-E port. The iButton is found on the following boards or modules
for each of these software versions: