EPIQ 5 & 7 Training Manual
© 2016 Conquest Imaging
62
Click
System Test
Run
.
Click
Yes
. The system test results are displayed in the
System
Test
box.
Click
OK
.
To export the system test results:
Perform the System Test.
In the
System Test Results
area, click
Export Logs
.
Connect a USB flash drive to a USB connector and select that port
as your export destination.
In
Select Logs to Export
, click
Include Diagnostic Logs
. You may
export other logs also, but the system test results are displayed
in the Diagnostic Logs.
Check the
Status
area for a successful export and click
OK
.
Filter Cleaning
Philips recommends once yearly cleaning at minimum and more often
in dusty environment. Also, you should think of this as one of these
very first “Is it plugged in?” troubleshooting steps. The system shuts
down if it overheats. The first and simplest probable cause is lack of
air flow.
Filter Cleaning Procedure
System power must be turned off.
Locate and slide filter out of left side.
Replace with spare filter and clean the filter first by vacuum,
wash with soap & water and dry. If a spare is not available
remember that the filter must be dry before replacement. A
damp filter could damage the system!
Reset the system’s “filter last cleaned date” as outlined below.