EasyScopeIII for DS60M10 User Guide
Version 1.3
Document Reference No.: CP_000037 Clearance No.: CP#027
Copyright © Connective Peripherals Pte Ltd
33
Test Mode
In test mode the EasyScope does not acquire real data but provides simulated data from a counter. This
allows the communication with the PC to be verified. The display observed in test mode should always be
a saw tooth type. The wave is at a frequency of 30MHz.
Figure 24- Test Mode
Test mode must be disabled to acquire real data.
Math Functions
The Math Function menu has three options as shown in the figure below.
A + B
This function adds the values of Channels A and B and displays the result waveform as a sum of these
two waveforms.
A - B
This Function subtracts the values of Channels B from Channel A and displays the result as a difference of
these two waveforms.
A x B
This function multiplies the values of Channels A and B and displays the result waveform as a product of
these two waveforms.
Important: Each of the resulting signals (A + B, A - B, A x B) has a corresponding volts/div
value. This value is displayed below the volts/div values for channel A and channel B.