5.3.8.2.1
Sensor Calibration
When calibrating with the
Sensor Calibration
method, each Twistlock will be affected individually. For each Twistlock with sensor you will
have to specify the according sensor coefficient, the Twistlock diameter and the Young’s modulus of the Twistlock steel. Then click on
Calibrate the Twistlock
and the Software will automatically calculate the overall Twistlock coefficient and store it in the interrogator. It is
used by the interrogator to calculate the weight on each Twistlock.
ADVICE!
The information about the Young’s modulus usually has large deviations.
Therefore this calibration method is the most inexact.