![Comdronic AC6-HP Скачать руководство пользователя страница 18](http://html.mh-extra.com/html/comdronic/ac6-hp/ac6-hp_instructions-manual_2639778018.webp)
18
7.
PcomPRO Project Application Software
Further options for saving valve data are available if PcomPRO Project Application Software is
purchased with the AC6 unit. PcomPRO gives the user the option of preparing a project on a PC
prior to commissioning the system. The data from the commissioning process can be saved to the
AC6 for later use.
8. Spares
and
Accessories
Replacement hose sets.
Binder-style adaptors.
Mechseal-style adaptors.
Industry-standard insertion test points.
Temperature and pressure gauges.
Replacement strainers.
9. Technical
Specification
Measurement Range / Accuracy
Model:
AC6-SP
AC6-HP
Accuracy:
+/- 1% of reading or +/- 0.1 kPa,
whichever is greater
+/- 1% of reading or +/- 0.2 kPa,
whichever is greater
Max Allowable
Static Pressure:
10 Bar
20 Bar
Measurement
Range:
0.5 kPa to 200 kPa
2 kPa to 600 kPa
Effective Operating Time
20 hours with standard Alkaline PP3 battery
Calibration
Standard factory calibration is carried out in controlled conditions using calibrators traceable to
National Standards. It is recommended that AC6 units are recalibrated at least once per year. This
recommended period is not fixed. Should the AC6 unit be used on systems with dirty water or a high
concentration of cleaning agents, treatments or the user suspects that readings are not perfect then
recalibrate more often.
Valve Database
The performance characteristics of over 4,500 valves – from 70 manufacturers – are pre-
programmed into the AC6.
Valve data is automatically updated whenever an AC6 meter is serviced by Comdronic Ltd.