11.5 ANALYZER
CONFIGURATION
THROUGH
THE
SERIAL
INTERFACE
The analyzer can use the serial interface to set parameters for weighing test
and for analyzer configuration. The parameters are set by sending a line of
data that has the description and the setting desired.
It is possible to create a text file that contains all the data necessary to set he
test conditions for a particular test (or analyzer configuration) and send this
to the analyzer. The commands are not case sensitive, i.e. parity=0 and
PARITY=0 are both accepted.
11.5.1
TEST PARAMETERS
The parameters that control a test are listed below. To change a test
parameter it is necessary to send a command for the parameter to modify
along with the value to change.
For example to change the test parameters to use a single heat and set the
temperature to 115ºC send 2 commands.
Heat=0<cr><lf>
Single Temp=115<cr><lf>
All commands are in English regardless of the language selected for the
analyzer operation. The commands can be sent in both upper case letters
and lower case letter, i.e. “!KT” or “!kt” are both acceptable.
Moisture Result=1;
#0=%M, 1=%S, 2=%Atro_M, 3=%Atro_S default=0;
Single Temp=100;
#050C to 160C,
default=100;
Interval Time=20;
#05 sec to 99 sec
default=15;
Stable Range=0.010;
#grams default=0.010
Stable Time=10;
#seconds, 01-99
default=02;
Start mode=0;
#0=Manual 1=Auto
default=0;
Care must be used as some commands depend upon others to be effective,
For example sending a command for setting the heat to use a ramp but not
sending a temperature or time to use may cause incorrect operation if they
are not set to reasonable values.