Page 17 of 60
6.1.1 MANUALLY SETTING TEST PROCEDURES
Press the [Test/ ] key to set the test procedure. During the process use the
numeric entry procedure described in section 4.1 to enter or change a
setting.
Pressing [Print/
] at any time during this procedure will print a list of the
current settings. See 6.1.2
Pressing [
0/T
/ Esc] will always escape back to normal weighing without
a change to the parameter being displayed. Any parameter changed before
this time will be accepted.
PROCEDURE
Press [Test/ ] to show “rESULt” with the current setting displayed,
Options are “Moisture”, “Solid”, “ATRO Moist”, “ATRO Solid”
Use arrow keys, [Dspl/ ] and [Test/ ], to change and press [Setup/Enter]
to continue.
“Single” Selected
Show “Single” “123 C” with the left digit flashing. The user can use the
[PST/
] key to select another digit or the [Dspl/ ] and [Test/ ] to set the
temperature, Range is 050 C to 160 C
Interval Setting
The next setting will be to set the time interval the analyzer uses to determine
results and to print the current result. The analyzer will compute the percent
ATRO MOIST
result
MOISTURE
result
SINGLE
ºC
123