Page 16 of 60
Interval
Set the time interval at which the results are computed and printed. Range is
1 to 99 seconds.
End Point Determination (manual stop active in all tests)
The analyzer ends a test when the results are stable within a value set by the
user. The test will stop when the weight changes less than a preset amount
in a preset time. The default setting is 0.002g in 15 seconds. The user can
set other values depending upon model.
Start Criteria
•
Start Manually when the [Start] key is pressed
•
Start Automatically when cover lowered.
Delays 5 seconds after cover is down before beginning.
Print settings
To set if the analyzer should print and store results and when printing if the
compete test results or only the summary is printed and stored in the
analyzer’s internal memory.
Default settings
% Moisture
Single Temperature = 110ºC
Rapid = Off
Interval Time = 5 seconds
End Point Determination = Time and stable result
Time 60 minutes, and
Stable = 0.002g/15 seconds
Start Criteria = manual start
Print Test = on
Print Format = Complete