Page 26 of 60
The Alternate displays in order are:
Mass
% Moisture
% Solid
% Atro Moisture
% Atro Solid
During the test the user can stop the test at any time by pressing the [Start]
key again. Otherwise it continues until the stop condition is met.
When the test has stopped the beeper will sound 3 short beeps (if enabled,
see sec. 10.1.2) and the display will show the end result:
This display shows the test was automatically stopped as stability was
reached in 12:44 minutes. If it had stopped at the end of the time period it
would have TSTOP on the top line and if manually stopped it would have
MSTOP.
At this time the final results are displayed. These results will include elapsed
time of the test and final % result.
Display other data by pressing the [Dspl/ ] key to show the initial mass and
final mass.
The final values for the results can be printed again to the Serial interface
by pressing the [Print/
] key. See section 8.
To return to normal weighing press the [
0/T
/ Esc] key.
12:44 ASTOP
76.47
%
START MASS
5.670
g
END MASS
1.334
g