moisture or solid at the time interval desired. The new result will be displayed
and printed as described below.
Display “Interval” “Sec XX” The user can change the value using the arrow
keys. Range is 01 to 99 Seconds. Typical results will be computed every 5-
15 seconds.
Enter the conditions to be used to determine if the results are stable. The
amount of weight change to be used to determine if the test is finished,
followed by the time to be used for the weight to change. For example 0.006g
in 15 seconds would be a typical setting. Longer times or smaller weight
changes could be used to ensure better stability or shorter times or larger
weight changes to ensure a faster determination. For example 0.006g in 5
seconds will end a test quicker (but possibly less accurately ) than 0.002g in
15 seconds. The user sets the weight first, then the time.
Display shows “ STAB RANGE” “ XXX”. Set the range for the weight
change. Set from 0.001 to 0.099.
Start Method
The last setting is the Start method. The analyzer will either start a test
automatically whenever the cover is lowered (after a 5 second delay to allow
stability), or manually when the [Start] key is pressed.
Display will show “START”
Options are “Manual” or “Auto”
INTERVAL
se[
[
[
[10
STOP RANGE
02
STOP TIME
sec15
MANUAL
start