background image

13

9.  

 SERVICE AND MAINTENANCE

 

9.1. Preventative maintenance checks should be carried as per the following table.

 

Page 13 of 24                                                                                                                                 M7819 Issue 4.2 

9.  SERVICE AND MAINTENANCE. 
 

9.1.  Preventative maintenance checks should be carried as per the following table. 
 

TABLE:-Preventative maintenance checks and service (PMCS) 

Item No 

Item to Check 
Service 

Procedure 

Not Fully Mission 
Capable If: 

1. 

Cover 

a.  Inspect for loose/missing parts and 

change physical damage. 

Parts such as screws 
are loose/missing or 
damage will prevents 
safe usage. 

 

 

b.  Check cover seal is properly fitted 

and not distorted. 

Cover seal is badly 
fitted or loose and 
leakage of steam or 
water is possible. 

 

 

c. 

Inspect for dirt, grease, oil or food 
debris. 

 

 

 

d.  Inspect for signs of corrosion. 

 

 

 

e.  Check steam vent hole is clear. 

 

Vent hole blocked. 

2. 

Inner Container 

a.  Check inner container is fitted and 

can be easily lifted out of the outer 
container 

Container is missing. 
Fold-down wire lifting 
handle is missing or 
broken. 

 

 

b.  Inspect for physical damage such 

as holes or large dents. 

Container has damage 
preventing safe usage. 

 

 

c. 

Inspect for dirt, grease, oil or food 
debris. 

 

 

 

d.  Inspect for signs of corrosion. 

 

 

3. 

Outer Container 

a.  Inspect for physical damage such 

as holes or large dents 

Container has damage 
preventing safe usage. 

 

 

b.  Inspect for dirt, grease, oil or food 

debris. 

 

 

 

c. 

Inspect for signs of corrosion. 
 

 

4. 

Main case 

a.  Inspect for loose / missing parts 

and physical damage. 

Parts such as screws 
are loose/missing or 
damage will prevents 
safe usage. 

 

 

b.  Check case seal is properly fitted 

and not distorted. 

Case seal is badly fitted 
or loose and leakage of 
steam or water is 
possible. 

 

 

c. 

Inspect for dirt, grease, oil or food 
debris. 

 

 

 

d.  Inspect of sign of corrosion. 

 

 

5. 

Latch 

a.  Check latch is secured to the main 

case and has no missing or broken 
parts. 

Latch has loose or 
missing screws. Any 
part is broken loose or 
missing. 

 

 

b.  Check lock and unlock operation is 

correct 

Latch will not lock / 
unlock. 

6. 

Tap 

a.  Check tap is secured to the main 

case and has no missing or broken 
parts. 

Tap has loose or 
missing screws. Any 
part is broken, loose or 
missing. 

 

 

b.  Check tap can be easily operated 

over its full range without binding. 

Tap dose not move 
freely or is jammed in 
one position. 

 
 
 
 

Содержание Electrothermal RAK30 HW & R

Страница 1: ...Heater Water Ration HW R RAK30 One shot boiling unit N S N 7310 99 362 3482 Instruction Manual M7819 Version 4 4...

Страница 2: ...l Support 21 Section 12 Notes 22 The copyright of this instruction manual is the property of Cole Parmer Limited This instruction manual is supplied by Cole Parmer Limited on the express understanding...

Страница 3: ...of water from 20 C in just over 30 minutes when switched to the HI setting After boiling the water will cool slowly until a temperature approaching 86 C 3 C is reached The HW R will then continue to o...

Страница 4: ...n normal domestic waste Caution Hot surface Refer to Instruction Manual Recyclable Packaging Material 4 2 SYMBOLS AND USING THIS INSTRUCTION BOOK 2 1 Throughout this Instruction book the following sym...

Страница 5: ...tion afforded by the equipment may not be preserved and the guarantee invalidated 3 SAFETY INFORMATION This product has been designed for safe operation when used as detailed in accordance with the ma...

Страница 6: ...em No Description Qty 1 Instruction Book 1 2 RAK3 0 HW R unit 1 For future reference please record your products Serial and Model Numbers Serial Number Model Number Page 6 of 24 M7819 Issue 4 2 4 UNPA...

Страница 7: ...e relevant vehicle installation data The three studs 5 16 x 10 UNC thread provide on the HW R are for location only and not for sole mounting purpose The recommended method for mounting of the HW R is...

Страница 8: ...ertificate are available from your distributor of Bibby Scientific products or you may copy and enlarge from Appendix A of this instruction book Bibby Scientific s Electrothermal branded product range...

Страница 9: ...nt components of the product 7 PRODUCT OPERATION 7 1 The RAK30 HW R has been designed for easy operation The illustrations below show detailed layouts of this product 1 Main Case Assembly 2 Outer Cont...

Страница 10: ...cle completed Heater is off Water is in process of cooling down to approximately 86 C This takes approximately 1 hour 4 Power On Boil Cycle Heater On LED On Green LED On Yellow LED On Yellow One shot...

Страница 11: ...ower On Boil Cycle Heater On LED On Green LED Off LED On Yellow Water is heating towards set point of 68 C 2 Power On Boil Cycle Heater On LED On Green LED Off LED Off Water is at 68 C set point Heate...

Страница 12: ...ed by temperature sensor using thermal feedback Heater Safety Cut out Boil dry temperature sensor operating at 172 C Non repairable thermal fuse set operating at 175 C Performance Data Heating water o...

Страница 13: ...holes or large dents Container has damage preventing safe usage c Inspect for dirt grease oil or food debris d Inspect for signs of corrosion 3 Outer Container a Inspect for physical damage such as h...

Страница 14: ...Inner Container a Check food has been removed b Inspect for dirt grease oil or food debris 12 Outer Container a Check water has been removed b Inspect for dirt grease oil or food debris 13 Main case...

Страница 15: ...to the surface finish DISHWASHING COMPOUND Discard hand dishwashing compound heated water solution inn accordance with the disposal instructions NOTE With the HW R mounted in the host vehicle it is no...

Страница 16: ...over perform the procedure in sub paragraph 5 2 16 9 2 15 Return the HW R to normal service Cleaning the Cover 9 2 2 Remove any loose food from the inner and outer containers by pre scraping 9 2 3 Fil...

Страница 17: ...are to avoid splashing the control panel 9 2 27 Allow main case to air dry Page 17 of 24 M7819 Issue 4 2 9 2 16 To prevent the ingress of liquid into the interior of the cover block off the small stea...

Страница 18: ...ining plate using an easily removable means e g tapered wooden peg of suitable size 18 Page 18 of 24 M7819 Issue 4 2 Sanitising the HW R unit 9 2 28 Fill the outer container with clean potable water u...

Страница 19: ...tap drain off the food service disinfectant heated water solution from the outer container 9 2 43 Half fill the outer container with clean potable water and rinse off all traces of food service disin...

Страница 20: ...f the HW R to cool Refill with fresh potable water 4 Electronics over temperature Switch off the HW R to cool If fault is repeated have HW R serviced Fault most likely in ambient conditions over 35 C...

Страница 21: ...al sales specialist or email cpspares coleparmer com to enquire about available spares Please visit www electrothermal com for a full list of available accessories 11 CUSTOMER AND TECHNICAL SUPPORT Fo...

Страница 22: ...12 NOTES 22...

Страница 23: ...23 12 NOTES...

Страница 24: ...Cole Parmer Ltd Beacon Road Stone Staffordshire ST15 0SA United Kingdom Tel 44 0 1785 812121 Email cpinfo coleparmer com Web www electrothermal com...

Отзывы: