Correct position of the fan
Reconnect the fan to the Main board and check the air flow direction.
- Dusty radiator.
In case the fans work correctly, but the air flow is not enough, it is possible that the
radiator externally is obstructed by dust, blocking the passage of air. To solve the
problem open the equipment and clean the radiator either with compressed air or remove
both fans for better Access and use a vacuum cleaner.
- Faulty applicator.
Check the real time temperatures of each plate (explained in previous chart). If the
values are not similar could be possible that some of the peltier elements is faulty and
can not reach the set temperature. Proceed replacing the applicator.
3.2.7 - Applicator temperature does not decrease
In case that one of the hand pieces does not reach the programmed temperature, but the other one
does a few minutes after start, exchange the applicators to determine if the fault is in the applicator or
in the channel. If the problem persists in the same hand piece proceed to replace it, but in case is in
same channel check the operation of the power source.
Open the equipment`s electic cabinet and measure the output voltage of the power source, supplying
the faulty applicator (SPV-300W-12V).
Содержание Cooltech
Страница 1: ...Cooltech SERVICE MANUAL Service Manual Cooltech V1 1 ...
Страница 5: ... A4 01454 x2 A4 01453 A4 01454 A4 01455 A4 01455 A4 01455 x2 x2 x2 ...
Страница 7: ...Packing layout A4 02266 A4 01389 x2 A4 01591 A4 01391 A4 01415 ...
Страница 8: ......
Страница 12: ...1 1 2 Vacuum system ...
Страница 37: ...Power supply mounting screws For better access to the screws ...
Страница 50: ...Power supply screws For better access to the screws ...