Remove the covers to access the
connector
Pin 4 and 5 of the lower connector
3.1.7 - Applicator temperature error
Problem in some temperature sensor or difference in temperature between the plates of a hand
piece. The treatment will be cancelled in case of having two faulty sensors in the same applicator, or
continue working if it is only one.
Whichever of this errors appears, it is recommended to replace the handpiece, although it is still
working.
There are 5 possible temperature reasons for this alarm, to find it out, it is necessary to access the
code screen, enter 123789 and press enter. The last 6 temperature alarms will appear, starting with
the newest (right) and the oldest (left). Each number has the following meaning:
0 :
No alarm
1 : NTC sensor short circuit
2 : NTC sensor open circuit
Содержание Cooltech
Страница 1: ...Cooltech SERVICE MANUAL Service Manual Cooltech V1 1 ...
Страница 5: ... A4 01454 x2 A4 01453 A4 01454 A4 01455 A4 01455 A4 01455 x2 x2 x2 ...
Страница 7: ...Packing layout A4 02266 A4 01389 x2 A4 01591 A4 01391 A4 01415 ...
Страница 8: ......
Страница 12: ...1 1 2 Vacuum system ...
Страница 37: ...Power supply mounting screws For better access to the screws ...
Страница 50: ...Power supply screws For better access to the screws ...