3.2.10 - Suction losses
If in case that the equipment reaches the programmed vacuum, but does not maintain it, most
probably is caused by an air leaks.In order to detect it follow the procedure step by step.
Vacuum circuit
The first step is to use the silicone pad or balloon in order to detect if the problem is caused by the
placement of the hand piece on the body or the coolpad is misplaced.
Step 1